- 1 PIEZOSURGERY® GP unit
- 1 torque wrench K8
- 1 carrying case
Handpiece and inserts not included
Now this is as easy as it gets. If you are looking for a device which provides simplicity and clarity in every detail – here you are: 100% simple handling for utmost treatment security.
Materials specially selected for easy cleaning, disinfection and sterilization. Cost-effective standard parts for perfect economy.
You call that perfect for daily use. We call it PIEZOSURGERY® GP.
With over 18 years of successful clinical research, the original PIEZOSURGERY® method is the only one supported by more than 250 clinical and scientific studies.
When it comes to cutting bone, you can of course use traditional burs and saws. They do cut bone, too – but they do not differentiate: any soft tissue getting in their way will also be cut.
The patented ultrasonic microvibrations of the original PIEZOSURGERY® technique cut bone – and nothing else. No soft tissue is damaged, which allows you to work with a precision that facilitates not only surgery itself, but reduces postoperative discomfort for your patients at the same time.
PIEZOSURGERY ® provides micrometric cuts for minimally invasive surgeries with maximum surgical precision and intra-operative tactile sensation.
Selective cuts protects any kind of soft tissue. Nerves, vessels and membranes will not be injured while cutting bone. Thus PIEZOSURGERY® offers maximum safety for surgeons and patients.
PIEZOSURGERY® offers maximum intra-operative visibility. The cavitation effect of the ultrasonic movements lead to a blood-free surgical site.
Comparative studies have demonstrated both the clinical and histological advantages of the PIEZOSURGERY® device.
Gleizal A, Li S, Pialat JB, Béziat JL.
Transcriptional expression of calvarial bone after treatment with low-intensity ultrasound: An in vitro study.
Ultrasound Med Biol. 2006; 32(10):1569-1574
The touch keyboard of the PIEZOSURGERY® white can be protected with a dedicated, individually packed, sterile transparent foil. Thanks to these invisible shields, no dirt, scratches or fingerprints will affect your keyboard.
Providing the optimal ratio between power and security is one of the key success factors of every surgery. Thanks to its intelligent electronic feedback system the original mectron PIEZOSURGERY® technology provides the maximum of power and achieves perfect cutting efficacy in every situation – for surgeries which are time-efficient, secure and successful.
Device compliant to Dir. 93/42/CEE: | Class II a |
Classification as per EN 60601-1: | I Applied part type B (handpiece, insert) IP 20 (device) IP X8 (foot pedal) |
Device for intermittent operation: | 60sec. ON - 30sec. OFF with irrigation 30sec. ON - 120sec. OFF without irrigation (ENDO,PERIO) |
Power supply voltage: | 100-240 Vac 50/60 Hz |
Max. power absorbed: | 120 VA |
Fuses: | Type 5 x 20 mm T 2AL, 250V |
Operating frequency: | Automatic scan From 24 KHz to 36 KHz |
Power types: | ENDO PERIO SPECIAL CANCELLOUS CORTICAL IMPLANT |
Peristaltic pump capacity: | Adjustable on the touch screen: ENDO / PERIO - 7 flow levels: from 0 to 6 (from 0 to approximately 75ml/min) SPECIAL / CANCELLOUS / CORTICAL / IMPLANT 6 capacity levels: from 1 to 6 (from 8 to approximately 75ml/min) |
LED system of the handpiece (optional): | LED system of the handpiece switches on as soon as the device starts to operate and switches off 3 seconds after the foot pedal is released |
Protections of the APC circuit: | No handpiece detected Cord interruption Insert not tightened correctly or broken |
Operating conditions: | from +10°C to +40°C Relative humidity from 30% to 75% Pressure of air P: 800hPa/1060hPa |
Transport and storage conditions: | from -10°C to +70°C Relative humidity from 10% to 90% Pressure of air P: 500hPa/1060hPa |
Peristaltic pump tube: | It is advisable not to exceed 8 sterilization cycles |
Weight | 3,2 Kg |
Size (L -W - H) | 300 x 250 x 95 mm |
Warranty | device 2 years; handpiece and handpiece cord 1 year |